| Input |
Strain Gauge Type Load Cell |
| Excitation Voltage |
DC5V,2.5V Current:Maximum 20mA
(By inside switch setting) |
| Sensitivity adjustment range |
0.5 to 3.0mV/V Can be converted to 0 to 1999 |
| Zero balancing |
Approx. ±3.0mV/V |
| Calibration |
1mV/V 1point |
| Display |
Range |
0 to 1999 |
| Decimal point |
Selectable(By inside switch) |
| Indicator |
Character:14.2mm
LED(Red) |
| Display timing |
Approx. 3times/sec. |
| Analog voltage output |
Full scale(0 to ±1999) Approx. 0 to ±2V
Load resistance: >1kohms |
| Comparator |
Point |
1Point(Selectable Hi/Lo Limit by inside switch)
|
| Setting |
Panel side Trimmerposition meter |
| Output |
Relay point of contact output(1T) |
| Capacity |
AC125V 0.5A(Resistance load)
|
| Movement |
Relay movement LED(lightly red) |
| Nonlinearity |
0.1%±1digit |
| Drift |
0.01%F.S./ºC
(Excitation Voltage 5V at 1mV/V Input) |
| Gain drift |
0.01%F.S./ºC |
| Environment |
Temperature |
-10ºC to 40ºC
|
| Humidity |
<80%RH(Non-condensing) |
| Power supply |
AC100V±10% 50/60Hz
Approx. 5VA
DC12V±10% Approx. 100mA |
| Dimension(W×H×D) |
Approx. 96mm×48mm×120mm(excluding tang) |
| Weight |
Approx. 400g |