A range of input amplifier cards, which can be directly connected to sensors measuring acceleration, temperature, revolutions, and distortion, or a CAN bus packet data recording card can be selected and used to construct a data recording system with a maximum of 64 channels (analog). With no need to carry an external amplifier and data verification equipment to the site, data can be easily verified in a range of environments, and then be swiftly transferred to a computer or analysis equipment. There are many possible recording methods including an AIT tape enabling prolonged recording or a built-in memory, MO, PC card or real time recording to a PC hard disk
Main features
  • As well as having a built in signal conditioner, its small A4 size removes limitations in terms of installation sites.
  • In addition to the various input/output amplifier cards, various recording media (memory, AIT, MO, PC card) are available.
  • Depending on phenomenon, a sampling frequency ranging from 1Hz to 200kHz, can be setup. It can be used in a range of conditions from prolonged data recording to recording of outbreak phenomenon when used in combination with recording media.
  • Setup, control, monitoring, and data transmission can be controlled from a Note PC (SCSI interface) using attached software.
  • A Color liquid crystal remote control is available for environments where PC's cannot be installed such as in automobiles.
  • Recording of CAN data, synchronized with an analog signal is also possible.
  • Software for the GX-1, which either directly controls recording, or imports the recorded data file and performs various processing tasks, is being offered by a number of venders.
Main specifications
Main Unit
Signal conditioner slot 8
Recording device slot 1(for AIT,MO or PC card drive selectable)
Memory 256MB max.
Interface ±2, 5, 10V
Contact-point control START, STOP, EVENT, REC/STANDBY
Expansion bus connector 1(common)
DIGITAL CONTROL connector 1(IRIG-B time code,external sampling clock input)
Monitor channel
analog output
1
Microphone jack 1
Speaker/earphone jack 1 each
Sampling frequencies [ External clock ]
Up to the maximum sampling frequency of each signal conditioner.

[ Internal clock ]
1/2/5/10/20/50/100/200/500/1k/2k/5k/10k/20k/50k/100k/200k Hz

[ Multi sampling ]
10:1
Note:The value [sampling-frequency × number-of- channels ] must not exceed 3,200kHz (in Memory type) and 1,500kHz (in ATI type) up to 32 channels.
Power 11V-30VDC, Approx. 45W (Memory-type,all channels DC signal conditioner installed)
Built-in AC adapter(separate adaptors for 100V and 200V)
Regulations:FCC part 15 class A,CE
Battery(Option)
Environment Temperature Operating range:0ºC to 40ºC(Memory type), 5ºC to 35ºC(AIT type)
Storage range:-20ºC to 60ºC
Humidity 20% to 80%RH(Non-condensing)
OS Windows XP/Windows 2000
Vibration MIL-STD-810D Figure 514.3-1, 2, 3 (Memory type)
MIL-STD-810C Figure 514.2-6 V Curve(1.5G,Vertical)(AIT type)
Dimension(W×H×D) Approx. 300mm×85mm×200mm
Approx. 11-13-16"×3-3/8"×7-7/8"
Weight Approx. 5kg/11.02lbs(Memory type)
Approx. 5.75Kg(AIT type)
(including 8 DC input amps and an AC adaptor)
Accessory AC power cable,3P-2P adaptor,DC power cable,AC adaptor,Earphone,Microphone,Ferrite core&cable tie 1 each,Ferrite core,Installation disk,Handle×2,Screws&washers for handle 4each,Hexagonal wrench,AIT cassette(AIT type only),Cleaning tape(AIT type only),MO disk(MO type only),Instruction manual(this manual)
Expansion Unit
■Dynamic Stain Input Amp AR-GXST
Dynamic Stain Input Amp AR-GXST Number of channel 2
Input format Balanced difference input
Input coupling DC
Input impedance 1Mohms
Gauge ratio 2.0
Excitation voltage DC2V,5V within ±0.2%
Has remote sensing.
Range ±200, 500, 1000, 2000, 5000, 10000µST
Frequency response 20Hz to 20kHz(-3dB)
Balance adjustment method  Electronic
Low pass filter 25/50/250/500/2.5k/5k/24k Hz,
PASS(-48dB/oct. Butterworth)
A/D converter 16bits
Sampling frequency 1Hz to 200kHz
Power consumption Approx. 3.4W
Input connector HR25-7P-8P (cable side) from Hirose Electric Co.,Ltd.
Input cable NDIS conversion cable(option)
With remote sense:CL-GXST2RS
Without remote sense:CL-GXST2
■Thermocouple Input Amp AR-GXTCK/J
Thermocouple Input Amp AR-GXTCK/J Number of channel 2
Applicable thermocouple K(AR-GXTCK),J(AR-GXTCJ)
Zero point compensation Electronic
Range AR-GXTCK -50ºC to 300ºC, -50ºC to 600ºC, -50ºC to 1200ºC
AR-GXTCJ -50ºC to 150ºC, -50ºC to 300ºC, -50 to 600ºC
Low pass filter 5Hz/10Hz/20Hz/PASS
A/D converter 16-bit, successive approximation type
Sampling frequency 1Hz to 200kHz
Power consumption Approx. 1.2W
Input connector GIM-K1(AR-GXTCK) from Omega
GIM-J1 (AR-GXTCJ) from Omega
■Charge Amp AR-GXCH
Charge Amp AR-GXCH Number of channel 2
Electric charge sensitivity 0.1pC/G to 999pC/G
Range 1G/5G/10G/50G/100G/500G(100G and the 500G range cannot be used when the electric charge sensitivity is 100pC/G or higher)
(1G and the 5G range cannot be used when the electric charge sensitivity is lower than 1pC/G)
Frequency range 1Hz to 30kHz(-3dB)
Low pass filter 100/200/500/1k/2k/5k/10k Hz(-48dB/oct.)
A/D converter 16bits
Sampling frequency 1Hz to 200kHz
Power consumption Approx. 2.16W
Input connector MICRO DOT 10-32UNF
■DC/ICP Input Amp AR-GXPAT
DC/ICP Input Amp AR-GXPAT Number of channel 2
Input format Unbalanced
Input coupling DC(DC mode)
AC(PA mode)
Input impedance 100kohms
Input range ±0.5/1/2/5/10/20Vp(AC, DC mode)
±0.1/0.2/0.5/1/2/5/10/20Vp(PA mode)
Low pass filter Attached filter of 40% of sampling frequency
Filter type: Butterworth(-48dB/oct.)
Sampling frequency 1Hz to 200kHz
Sensor power supply 24V DC/4mA(PA mode)
Frequency response DC to 80kHz(DC mode)
0.1Hz to 80kHZ(AC/PA mode)
A/D converter 16bits
Input connector BNC
■Analog Output Amp AR-GXAO
Analog Output Amp AR-GXAO Number of channel 2
Input format Unbalanced
Input coupling DC
Output level ±1 to 5Vp(in 0.1V steps)
Zero shift 0 to ±5V(in 0.1V steps)
Frequency response DC to 80kHz
Low pass filter 2/4/8/20/40/80/200/400/800/2k/4k/5k/8k/10k/20k/40k/80k Hz
Filter type:Butterworth(-48dB/oct.)
D/A converter 16bits
Input connector BNC
Option
■Battery Pack(HP-30A)
Supply voltage 13.2V
Capacity 3.7Ah
Size NP-1 type
■Battery charger(KH-2S)
Power supply 100VAC(Automatic switching to 200V AC)
Number of slot forbattery pack 4slots (each 2packs rechargeable at the same time)
*Regulations:UL,CSA,CE
■Remote Controller(ER-GXRC)
Function Record,Playback(START,STOP,EVENT,REC)
*Using contact control
Expansion Unit(AU-GXEPIO)
Signal conditioner slots 8
Power 11VDC to 30VDC,
Internal AC adaptor
(separate adaptors for 100V and 200V)
Battery (optional, same as for main unit)
External dimensions(W×H×D) Approx. 300mm×85mm×200mm
Approx. 11-13/16"×3-3/8"×7-7/8"
(excluding protruding parts)
Weight Approx. 5kg(11.02lbs.)