| Type of Input Signal |
Strain gauge type transduce |
| Input Signal |
0.5 mV/V to 3.2 mV/V |
| ExcitationVoltage |
DC 10V, 5V, 2.5V selectable by switch |
| Maximum Current |
within 60mA
Up to 4 pcs. 350 ohm transducer can be connected.
(When using 5V or 2.5V excitation) |
| Adjustable Zero Offset |
±0.25mV/V adjustable by potentiometer(8rotations) |
| Current Output :Voltage Output |
Fixed 4.00mA at 0V |
| Adjustable Sensitivity |
At 10V excitation |
Adjustable 0.5 mV/V to 20mA,10V |
| At 5V excitation |
Adjustable 1mV/V to 20mA, 10V |
| At 2.5V excitation. |
Adjustable 1mV/V to 20mA, 5V |
| Output |
Voltage |
0 to ±10V (> 2 k ohms) |
| Current |
4 to 20mA(< 500 ohms) |
| Non-Linearity |
< 0.02% F.S
(At 10V excitation, adjusted to 1mV/V to 10V) |
| Frequency Bandwidth |
DC to 10 Hz(-3 dB) StandardVersion |
| Calibration Signal |
1 mV/V(added to input signal) |
| Error |
< ± 0.3% F.S |
| Stability |
|
| Zero Drift |
0.05 % F.S. / ºC |
| Sensitivity Drift |
0.03 % F.S. / ºC (At 10V excitation, adjusted to 1mV/V to 10V) |
| Noise |
< 1µVp-p(Converted to Input level)
(Voltage output from 0.1 to 10 Hz excluding spike noise)
(Gain x 1000, adjusted 1mV/V to 10V at excitation 10V) |
| Operation Temperature |
0 to 40ºC |
| Operating Humidity |
Max 80% RH(no condensation) |
| Power |
AC 85 to 132V Approx. 5VA
DC 20 to 30V Approx. 0.3 A
AC type or DC type has to be determined at time of order |
| Dimension(W×H×D) |
50mm×83mm×44mm excluding socket |
| Weight |
Approx 240g(AC model), 180g(DC model) |
| Accessories |
Socket(PF113A) |
1 |
| Spring for Lock |
1 |
| Small Screw Driver(for Adjustment) |
1 |
| InstructionmAnual |
1 |